CREATE REQUEST


Main advantages

  • flexibly adapts to any mounting situation
  • quick and easy change from compact to separate modification
  • powerful converter for many applications
  • robust and resistant cover of sensor and converter

Applications

  • water treatment
  • mechanical engineering
  • energy engineering
  • food industry
  • agriculture

Product photogallery

Practical installation


Technical data of control unit

MQI 99-C
Electric conductivity of medium ≥5 µS/cm, for demineralized water ≥20 µS/cm
Input resistance ≥10·10 Ohms
Accuracy of measurement 0,3 % of measured value in range of 10 ÷ 100 % Qmax
Filtration of measurement adjustable in multiple modes
Elimination of small flows adjustable by 0,1 %
Direction of flows Bi-directional measuring indicated by the sign
Zero flow automatic zero point setting
Data recording flow volume archiving, even during power supply failure
Statistics 5 minutes – hourly – daily – monthly for 4 months
Real time clock with calendar inc. leap years until 2099
Values displaying alphanumeric LCD, 2 × 16 characters, with backlight
Control keypad with four buttons
Programmable functions detection of empty pipe
Analog output (active) galvanically separated 4 (0) ÷ 20 mA, (0 ÷ 30 mA) / 300 Ohms
Impulse output (1× relay) flow volume (x·m³) per impulse
Binary output (3× relay) to order (non-inductive load) 30 V/2 A/DC^125 V/0,5 A/AC
Frequency output (open collector) to order nominal rating of 1 kHz/TTL
Communication output (data) to order RS 232 C / RS 485, selectable speed, protocol ELA
Power supply 85 – 260 VAC 50 / 60 Hz, 24 VDC, to order 9 – 36 VDC +10% / -15%, 10 VA
Enclosure IP 67 (NEMA 5)
Ambient temperature −20 °C to +50 °C
Dimensions (height × width × depth) ∅115 × – × 180 mm
Material of case casting ALSi10Mg/EN AC-47100, surface finish STANDARD-Elox DIN 17611 EO, E6/EV1-natural color
Weight 2,6 kg
Cable terminals 3 pcs PG 11

Basic dimensions


Technical data of flange sensors

Flange sensor variants design 1 design 2 design 3
Control principle DC pulse
Connecting cable standard 0,5 m, to order 16, 24, 32, 40, 48 m, cable can not be extended without consultation with producer
Construction completely welded carbon steel
Inner diameter DN 10 ÷ 1000
Mounting connection flange DIN, ANSI, BS
Enclosure IP 67 / NEMA 5 (IP 68 / NEMA 6)
Measuring tube material stainless steel
Sensor cover material carbon (stainless) steel
Flange material carbon (stainless) steel
Ground rings DN 10 ÷ DN 40 to order for installation in plastic pipe
Sensing electrode material stainless steel (titanium, hastelloy, platinum)
Lining material hard rubber soft rubber teflon PTFE
Maximum temperature of medium * 90 °C 90 °C 130 °C
Standard pressure load
Pressure range **
1,6 MPa
0,6 / 1,0 / 2,5 / 4,0 MPa
1,6 MPa
0,6 / 1,0 / 2,5 / 4,0 MPa
1,6 MPa
0,6 / 1,0 / 2,5 MPa

* Maximum temperature of measured medium consult with producer
** Construction of sensor and type of flange depends on the maximum desired pressure of medium