MQI 99-C |
| Electric conductivity of medium |
≥5 µS/cm, for demineralized water
≥20 µS/cm |
| Input resistance |
≥10·10 Ohms |
| Accuracy of measurement |
0,3 % of measured value in range of 10 ÷
100 % Qmax |
| Filtration of measurement |
adjustable in multiple modes |
| Elimination of small flows |
adjustable by 0,1 % |
| Direction of flows |
Bi-directional measuring indicated by the sign |
| Zero flow |
automatic zero point setting |
| Data recording |
flow volume archiving, even during power supply failure |
| Statistics |
5 minutes - hourly - daily - monthly for 4 months |
| Real time |
clock with calendar inc. leap years until 2099 |
| Values displaying |
alphanumeric LCD, 2 × 16 characters, with backlight |
| Control |
keypad with four buttons |
| Programmable functions |
detection of empty pipe |
| Analog output (active) |
galvanically separated 4 (0) ÷ 20 mA, (0 ÷ 30 mA) / 300
Ohms |
| Impulse output (1× relay) |
flow volume (x·m³) per impulse |
| Binary output (3× relay) |
to order (non-inductive load) 30 V/2 A/DC^125 V/0,5 A/AC |
| Frequency output (open collector) |
to order nominal rating of 1 kHz/TTL |
| Communication output (data) |
to order RS 232 C / RS 485, selectable speed, protocol
ELA |
| Power supply |
85 - 260 VAC 50 / 60 Hz, 24 VDC, to order 9 - 36 VDC
+10% / -15%, 10 VA |
| Enclosure |
IP 67 (NEMA 5) |
| Ambient temperature |
−20 °C to +50 °C |
| Dimensions (height × width × depth) |
∅115 × - × 180 mm |
| Material of case |
casting ALSi10Mg/EN AC-47100, surface finish
STANDARD-Elox DIN 17611 EO, E6/EV1-natural color |
| Weight |
2,6 kg |
| Cable terminals |
3 pcs PG 11 |